Right Heart Catheterization — Technique and Measurement

Definition

The measurement half of right heart catheterisation is a sequence of decisions — where the catheter sits, where zero is, which phase of the breath and of the cardiac cycle the pressure is read in, and which method measures output — and each one moves the number. A pressure recorded with the wrong zero, read against the wrong part of the waveform, or paired with an output from an inappropriate method is still a number, and it will still be acted on.

Key Concepts

Access, Position and Zero Reference

Cardiac Output Measurement

PAWP Technique

Respiratory Cycle and Procedural Conduct

Mixed Venous Saturation and Shunt Detection

Safety and Contraindications

Reference Values

Variable Formula Normal Clinical application
RAP Direct 2–6 mmHg (0–6 in sources/rhc-hf-jacchf-2024) RV failure, shock staging
RV pressure Direct Systolic 20–34; diastolic 0–6 mmHg RV failure; RVOT obstruction
PA pressure Direct Systolic 20–34; diastolic 8–12 mmHg Pulmonary hypertension; RV afterload
mPAP Direct 8–20 mmHg PH diagnosis (>20 mmHg)
PAWP Direct ≤13 mmHg LV filling pressure; PH classification
CO Direct 4–8 L/min Pump function
CI CO/BSA 2.5–4.0 L/min/m² Standardised pump function
PVR (mPAP − PAWP)/CO 0.3–2.0 WU Pulmonary vascular disease
TPG mPAP − PAWP ≤12 mmHg Pulmonary vascular disease; transplant candidacy
DPG dPAP − PAWP <7 mmHg Pulmonary artery remodelling marker
RAP:PAWP Direct <0.5 Biventricular failure, constrictive physiology
SVRI (MAP − RAP)/CI ~10–15 WU Shock differentiation
CPO (CO × MAP)/451 ~0.8–1.1 W Cardiogenic shock severity
RVSWI (CI/HR) × (mPAP − RAP) × 0.0136 ~5–10 g·m/m²/beat RV workload; LVAD risk
PAPI (sPAP − dPAP)/RAP >~0.9–1.0 RV pulsatility; cardiogenic shock

(sources/rhc-hf-ehj-2025, rating: high)

The normal ranges are not uniform across sources. The practical-guidance review gives RA 0–6, RV systolic 20–34 and diastolic 0–6, PA systolic 20–34 and diastolic 8–12, and PCWP 4–14 mmHg (sources/rhc-hf-jacchf-2024, rating: high). Two entries above differ from that set and are left visible rather than reconciled: RAP is given as 2–6 mmHg here against 0–6 there, and PAWP as ≤13 mmHg against a 4–14 mmHg range. The gaps reflect different reference populations rather than a dispute about physiology, but a right atrial pressure of 1 mmHg or a wedge pressure of 14 mmHg sits inside one range and outside the other — so a quoted "normal" should name its source.

Contradictions / Open Questions

Connections