Electrophysiological Characteristics of Septal Perforation During Left Bundle Branch Pacing

Authors, Journal, Affiliations, Type, DOI

Overview

Septal perforation is the complication every LBBP operator fears, but its electrophysiological signature had not been characterised. This study analyses 30 perforations among 212 successful implants (14.1%) and describes what the lead sees as it enters the LV cavity: unipolar impedance falls to ~405 Ω, the current of injury collapses from ~15 mV to ~1 mV, and the unfiltered unipolar electrogram changes shape in one of two ways. A QS pattern (type I, 67%) indicates complete perforation, with the helix free in the cavity and capture only at high output. An RS/rS pattern (type II, 33%) indicates partial perforation with part of the helix still in contact — which still captures in 80%, but at a high threshold. A unipolar impedance cut-off of <450 Ω diagnosed perforation with sensitivity 100%, specificity 96.6%, PPV 83.3% and NPV 100%. All 30 leads were withdrawn and reimplanted at a different site, with no VSD, septal haematoma, coronary injury, embolism, or dislodgement during follow-up. The paper's central argument is that perforation recognised at the table and managed by re-siting the lead is benign — but its 14.1% rate is 4–40 times higher than previously reported, and the diagnosis is made by the same electrical parameters being evaluated.

Keywords

Current of injury; Left bundle branch pacing; Pacing impedance; Septal perforation; Template beat; Unfiltered unipolar electrogram

Key Takeaways

Cohort and incidence

The impedance signature

Cut-off Sensitivity Specificity PPV NPV Accuracy
<500 Ω 100% 86.6% 55.5% 100% 88.5%
<450 Ω 100% 96.6% 83.3% 100% 97.13%
<400 Ω 46.6% 99.4% 93.3% 91.7% 91.8%

The current-of-injury signature

Two unfiltered unipolar electrogram patterns

Feature Type I — QS Type II — RS/rS
Incidence 20 (67%) 10 (33%)
Interpretation Complete perforation — helix in LV cavity Partial perforation — part of helix still in septal contact
Impedance 402.5 ± 20.4 Ω 411 ± 21.3 Ω
COI before → after 12.7 ± 9.3 → 0.9 ± 0.6 mV 19.7 ± 10.5 → 1.3 ± 0.9 mV
Capture during perforation 7/20 (35%), threshold 3.3 ± 0.4 V 8/10 (80%), threshold 2.8 ± 0.8 V

Management and outcome