Spherical Monopolar Irrigated RF Catheter — Ovine Ventricular Lesion Study (2019)

Authors, Journal, Affiliations, Type, DOI

Overview

The first ventricular lesion comparison for the Sphere-9 expandable spherical irrigated RF electrode — a 9-mm nitinol lattice that presents an effective surface area of about 275 mm² against 28 mm² for a conventional 3.5-mm tip, allowing roughly 170 W to be delivered at low current density. In 11 sheep, 20 lesions created with the sphere were compared with 25 created by a conventional irrigated catheter in healthy left ventricular endocardium: lesion volume was three times larger (1,707 vs 563 mm³) and depth roughly double (10.3 vs 5.5 mm), with no steam pops in the sphere group and one in the conventional group. The authors argue that 10-mm lesions would allow transmural ablation across almost any region of a preserved-thickness left ventricle, potentially reducing or removing the need for epicardial access — a claim that framed the subsequent clinical programme but has never been tested against epicardial ablation.

Keywords

Ablation; novel ablation catheter; spherical monopolar irrigated ablation catheter; ventricular tachycardia

Key Takeaways

Device and rationale

Lesion comparison

Parameter Spherical (n=20) Conventional irrigated (n=25) P
RF duration 91.7 ± 29.5 s 63.2 ± 6.5 s .01
Mean power 169 ± 28 W 30.2 ± 1.2 W <.001
Initial impedance 46.1 ± 2.3 Ω 98 ± 22 Ω <.001
Δ impedance 5.5 ± 1.9 Ω 12.5 ± 6.2 Ω .013
Mean temperature 55.3 ± 4.2 °C 41.2 ± 6.2 °C .001
Lesion depth 10.3 ± 2.9 mm 5.5 ± 2.4 mm .001
Lesion width 11.6 ± 4.2 mm 5.2 ± 1.8 mm .001
Lesion volume 1,707 ± 771 mm³ 563 ± 398 mm³ .001