Conduction System Pacing-Induced Ventricular Arrhythmia: Case Report
Authors, Journal, Affiliations, Type, DOI
- Authors: Serhat Kesriklioglu, Ahmet Taha Sahin, Mertcan Gezer, Mustafa Celik, Ahmet Lutfu Sertdemir, Enes Elvin Gul
- Journal: Annals of Noninvasive Electrocardiology 2026;31:e70190
- Affiliation: Department of Cardiology, School of Medicine, Necmettin Erbakan University, Konya, Türkiye
- Type: Case report
- DOI: 10.1111/anec.70190
Overview
This case report describes a 78-year-old woman with pacing-induced cardiomyopathy who underwent an upgrade from a dual-chamber pacemaker to LOT-CRT. LBBAP produced only deep septal pacing (DSP) with a prolonged V5 R-wave peak time, so a coronary sinus LV lead was added. The procedure acutely narrowed the paced QRS from 185 to 132 ms, but was accompanied by marked QT prolongation and T-wave inversions. On the first post-implant day she developed ventricular fibrillation triggered by premature ventricular complexes, without a reversible metabolic or ischaemic cause and with stable device parameters. She was treated with propranolol, remained arrhythmia-free for 10 days, and was not upgraded to an ICD because there were no further episodes. The report is a reminder that acute electrical resynchronisation does not guarantee electrical stability, and that early post-implant arrhythmic vulnerability may exist in severely remodelled ventricles.
Keywords
Arrhythmias; Cardiac resynchronization therapy; Case report; Left bundle branch area pacing; Ventricle
Key Takeaways
Case
- 78-year-old woman with a dual-chamber pacemaker implanted in 2020 for complete heart block; admitted for CRT upgrade because of progressive pacing-induced cardiomyopathy.
- NT-proBNP 19,032 pg/mL; LVEF 35%; severe mitral regurgitation; moderate tricuspid regurgitation.
- Baseline paced QRS 185 ms.
- LBBAP attempted with a Medtronic 3830 lumenless lead; multiple attempts achieved only deep septal pacing with V5 RWPT 104 ms.
- A coronary sinus LV lead was added (lateral branch) to complete LOT-CRT.
- Final paced QRS 132 ms with appropriate capture, but the ECG showed QT prolongation and T-wave inversions in precordial and inferior leads.
- Day 1 post-procedure: ventricular fibrillation requiring defibrillation.
- No reversible metabolic or ischaemic trigger; device interrogation showed appropriate sensing and capture, no lead dislodgement; echo showed no septal haematoma, no pericardial effusion, and no new regional wall motion abnormality.
- PVC morphology suggested the arrhythmia was triggered by the LBBAP lead rather than the LV lead.
- Propranolol 40 mg orally three times daily was started to shorten QT and suppress ventricular arrhythmia; other antiarrhythmics were avoided because of QT prolongation.
- No further arrhythmia during a 10-day hospital stay; ICD upgrade was not performed because there had been only one episode.
Discussion points
- LBBAP generally reduces transmural dispersion of repolarisation and has a lower proarrhythmic risk than BVP in the I-CLAS analysis, but proarrhythmia remains possible.
- Proposed mechanisms: shifts in depolarisation–repolarisation dynamics after restoration of synchrony; heterogeneous recovery of action potentials in remodeled myocardium; regional repolarisation dispersion from selective or partial fascicular capture.
- The early post-LBBAP period may be a window of increased arrhythmic vulnerability in severely compromised ventricles, independent of procedural success.
- Prior case reports cited: bundle branch re-entrant VT during LBBAP implantation (Ponnusamy and Vijayaraman 2021); monomorphic VT months after LBBP (Tanawuttiwat 2023); electrical storm with fascicular-like PVCs resolving after lead removal (Bhatt 2024); delayed septal perforation causing VT (Hsieh 2024).
Limitations of the document
- Single case report — no control group, no incidence estimate, and no ability to establish causality.
- Attribution of the arrhythmia to LBBAP is inferential, based on PVC morphology; the LV lead, the underlying cardiomyopathy, and the procedure itself cannot be excluded as contributors.
- No electrophysiological study or mapping was performed to confirm the arrhythmia mechanism.
- Short follow-up (10 days) — late recurrence and the appropriateness of withholding an ICD are unknown.
- QT prolongation and T-wave inversion are described but not quantified (no QTc values), limiting mechanistic interpretation.
- No genetic or fibrosis characterisation to explain individual susceptibility.
- Consent and publication: written informed consent obtained; authors report no conflicts of interest.
Key Concepts Mentioned
- concepts/Conduction-System-Pacing — the broader technique
- concepts/Left-Bundle-Branch-Area-Pacing — the pacing strategy
- concepts/LOT-CRT — the hybrid configuration used
- concepts/Pacing-Induced-Cardiomyopathy — the indication for upgrade
- concepts/Sudden-Cardiac-Death — the adverse event
Key Entities Mentioned
- Necmettin Erbakan University (Konya) — the reporting centre
- Medtronic — manufacturer of the 3830 lead