Electrophysiological characteristics of lead position–dependent electrogram uninterrupted transition during left bundle branch pacing

Authors, Journal, Affiliations, Type, DOI

Overview

The conventional way to confirm left bundle branch capture is interruptive: the operator stops screwing the lead, tests pacing output, reads the ECG and electrogram, then advances again. That stop-start approach means the continuous transition of the paced ECG and intracardiac electrogram as the lead crosses the septum is never observed — only sampled, with the risk that subtle transitions are overlooked or misinterpreted. This study used a continuous pacing-and-recording technique enabled by a rotatable connector to watch those transitions in real time through lead advancement, in 105 patients. The principal finding is that the ventricular current of injury (COI) — the unfiltered unipolar injury signal at the lead tip — changes predictably with radial septal depth, peaking within the interventricular septum, and that a discrete COI component may serve as a novel marker of selective LBBP. The practical claim is that continuous COI and Stim-V6RWPT monitoring tells the operator where the lead is while it is being advanced, which both simplifies implantation and may allow earlier detection of septal perforation.

Keywords

Left bundle branch pacing; Continuous recording technique; Unfiltered unipolar electrogram; Current of injury; Septal perforation

Key Takeaways

Background and Objectives

Methods

Definitions of LBB Capture — the two strict criteria

Dynamic paced QRS transition is the criterion for LBB capture diagnosis. LBB capture is confirmed by a paced QRS morphology showing right bundle branch conduction-delay pattern plus these criteria:

  1. If the paced QRS morphology changes with a transition in lead V1, and the Stim-V6RWPT for 2 adjacent paced QRS complexes abruptly decreases by ≥10 ms with a constant output (2 V/0.5 ms) during lead screwing, and remains shortest and constant at low and high outputs → confirms LVSP has changed to NSLBBP
  2. If the paced QRS morphology changes with a transition in lead V1, and an isoelectric interval and discrete component appear on the EGM with fixed Stim-V6RWPT → confirms NSLBBP has changed to SLBBP

Statistical Analysis

Bland-Altman statistics assessed interobserver agreement of the Stim-V6RWPT and COI amplitude measurements (reported in the Online Supplemental Results and Figure 1)

Results — Baseline Characteristics (Table 1)

Results — Pacing Modalities Achieved

Results — Lead Position–Dependent Parameters (Tables 2 and 3)

Procedure-related parameters (Table 2, n = 88):

Parameter Value
Incidence of PoLBB 66 (75.0%)
PoLBB-ventricular interval 24.7 ± 8.1 ms
PoLBB-V6RWPT 64.5 ± 7.1 ms
PoLBB amplitude 0.35 ± 0.34 mV
Abrupt shortening of Stim-V6RWPT 14.3 ± 6.1 ms
Initial LBB capture threshold 0.76 ± 0.56 V (at 0.5 ms)
Initial LVS capture threshold 1.32 ± 0.91 V
Initial R-wave amplitude 8.1 ± 3.9 mV
Initial impedance 783.6 ± 145.9 Ω
Final LBB capture threshold 0.61 ± 0.44 V
Final LVS capture threshold 0.66 ± 0.25 V
Final R-wave amplitude 15.1 ± 7.1 mV
Final impedance 743.1 ± 136.8 Ω
Lead depth 15.1 ± 2.6 mm
Perforation 3 (3.4%)

Position-dependent electrophysiology (Table 3, by pacing modality):