Antitachycardia Pacing at the His Bundle is Safer than Conventional Right Ventricular Antitachycardia Pacing in a Canine Myocardial Ischemic Injury Model

Authors, Journal, Affiliations, Type, DOI

Overview

This preclinical study tested whether antitachycardia pacing (ATP) delivered from a His-bundle lead is more effective and safer than conventional right ventricular (RV) apical ATP. Six mongrel dogs underwent left anterior descending artery occlusion followed by reperfusion, with pacing leads in the RV apex and His bundle; 83 VT episodes were induced and 128 ATP trials were delivered, crossing over between sites when the first failed. His ATP terminated VT at a similar rate to RV ATP (23 vs 25 successful trials), but caused significantly fewer episodes of VT acceleration (6 vs 14, p = 0.05) and degeneration into ventricular fibrillation (1 vs 7, p = 0.02). Epicardial sock mapping showed more complete myocardial capture from the His lead, which the authors attribute to Purkinje-mediated spread and a less curved activation wavefront. The study is the first test of conduction-system ATP and concludes that His ATP is safer, but not more effective, than RV ATP.

Keywords

His bundle pacing; antitachycardia pacing; ventricular tachycardia; VT acceleration; conduction system pacing

Key Takeaways

Methods

VT characteristics and ATP delivery

Efficacy

Safety

Myocardial capture and mechanism

Limitations of the document

Key Concepts Mentioned

Key Entities Mentioned

Wiki Pages Updated