AFFERA Ventricular Arrhythmia Ablation Registry (AVAAR) — 2025

Authors, Journal, Affiliations, Type, DOI

Overview

The largest systematic dataset for any pulsed-field ventricular ablation platform: every ventricular arrhythmia procedure performed with the AFFERA dual-energy lattice-tip catheter (Sphere-9) at all 18 European centres using the system, 126 patients, minimum 3-month follow-up. It is a safety registry as much as an efficacy study, and it reads like one. Major acute complications occurred in 6% and minor in 14%; 8 patients (6%) had major events between day 1 and 30 including 3 deaths, and the causes are instructive — coronary laceration at epicardial access, progressive heart failure in a patient with LVEF 15%, sepsis. Acute success was 84% and 72% remained free of their arrhythmia at 5.6 ± 3.7 months, with de novo procedures outperforming redo (80% vs 68%). The registry also documents the platform-specific phenomena that matter in practice: coronary spasm with epicardial and RVOT delivery, three cases of haemolysis without renal failure, a catheter-shaft insulation rupture that caused simultaneous VF and AF during RF, electromagnetic interference rendering mapping impossible in two left ventricular assist device patients, and one false tendon avulsed during catheter manipulation.

Keywords

VT ablation; pulsed field ablation; lattice-tip catheter; safety

Key Takeaways

Design and population

Acute complications (procedure to day 1)

Complications from day 1 to 1 month

Efficacy

Interpretation offered by the authors