Lead I R-Wave Peak Time for Conduction System Capture in LBBAP

Authors, Journal, Affiliations, Type, DOI

Overview

Lead V6 R-wave peak time is the workhorse criterion for left bundle branch capture, but its specificity collapses when V6 shows an rS morphology, and it varies with cycle length and respiration. This research letter externally validates lead I R-wave peak time as a substitute in 217 patients at one Spanish centre. Lead I outperformed lead V6 in every analysis: AUC 0.934 versus 0.807 in patients with a normal conduction system and 0.904 versus 0.895 in conduction system disease; Cohen's κ against the capture reference was 0.81 versus 0.59. The advantage was largest exactly where V6 fails — in patients with rS in V6, lead I kept 80% specificity while V6 fell to 10%. The study validates cut-offs proposed elsewhere rather than deriving new ones, and it is the first external test of those values.

Keywords

Left bundle branch pacing, conduction system capture, left septal pacing, R-wave peak time, lead I, lead V6, diagnostic accuracy, rS morphology

Key Takeaways

Rationale

Design

Population

Derived Optimal Cut-Offs

Group Lead AUC Cut-off Sensitivity Specificity
Normal conduction system Lead I 0.934 82 ms 94.7% 88.9%
Normal conduction system Lead V6 0.807 80 ms 90.7% 66.7%
Normal conduction system, V6 rS excluded Lead I 0.934 82 ms 93.2% 84.6%
Normal conduction system, V6 rS excluded Lead V6 0.915 80 ms 88.1% 92.3%
Conduction system disease Lead I 0.904 88 ms 95.7% 86.2%
Conduction system disease Lead V6 0.895 80.5 ms 85.9% 86.2%
Conduction system disease, V6 rS excluded Lead I 0.959 88 ms 94.4% 91.7%
Conduction system disease, V6 rS excluded Lead V6 0.959 85 ms 88.9% 95.8%

Overall Performance

The rS Subgroup — Where the Criterion Matters Most

Interpretation

Limitations of the document