Durability of Atrial Linear Lesion Sets with a Dual-Energy Lattice-Tip Catheter

Authors, Journal, Affiliations, Type, DOI

Overview

This study reports invasive remapping of left-atrial linear lesions created with the dual-energy lattice-tip (Sphere-9) catheter, which can toggle between radiofrequency and pulsed field energy. Of 252 patients treated, 39 (16%) returned for a repeat procedure, allowing 36 pulmonary veins and 71 linear lesions to be reassessed. Posterior-wall lesions created with pulsed field alone were completely durable — roof lines 6/6 and posterior box lesions 10/10 — as were all 7 cavotricuspid isthmus lines made with radiofrequency. Anterior lines (9/17, 53%) and mitral isthmus lines (7/13, 54%) were not, reconnecting at roughly the rate reported for conventional thermal ablation. The striking result is therefore narrow: adding PFA to the lattice tip has solved the posterior wall, where oesophageal proximity limited thermal energy, and has not solved the anterior wall or mitral isthmus, where the obstacle is myocardial thickness and anatomy rather than heat.

Keywords

Atrial fibrillation, catheter ablation, pulsed field ablation, radiofrequency ablation, lattice-tip catheter, dual-energy ablation, linear ablation, lesion durability, posterior wall isolation, mitral isthmus, cavotricuspid isthmus

Key Takeaways

Rationale

Design

Cohort

Durability Results

Lesion Durable / reassessed Durability Energy Median days to remap
Pulmonary veins 31/36 86% PFA only 199 (102–579)
Roof line 6/6 100% PFA only 564 (175–834)
Posterior box 10/10 100% PFA only 58 (35–175)
Anterior line 9/17 53% RF + PFA 46 (35–81)
Mitral isthmus 7/13 54% RF ± PFA 35 (29–67)
Cavotricuspid isthmus 7/7 100% RF only 43 (34–175)
All linear lesions 54/71 76% — —